(浙)浙江省地方计量技术规范JJF(浙)1074—2012————————————————————————————————空气微泄漏检测仪校准规范CalibrationSpecificationforAirLeakageTester2012-12-28发布2013-02-01实施————————————————————————————————浙江省质量技术监督局发布JJF(浙)1074--2012空气微泄漏检测仪校准规范CalibrationSpecificationforAirLeakageTester归口单位:浙江省质量技术监督局主要起草单位:浙江省计量科学研究院本规范委托主要起草单位负责解释JJF(浙)1074-2012JJF(浙)1074--2012本规范主要起草人:陈宇航(浙江省计量科学研究院)童敏之(浙江省计量科学研究院)蔡绯(浙江省计量科学研究院)JJF(浙)1074--2012I目录引言.......................................................................................................................(II)1范围............................................................................................................................32引用文件....................................................................................................................33术语............................................................................................................................33.1基准物.....................................................................................................................33.2泄漏量.....................................................................................................................33.3差压式检漏仪......................................................................................................34概述............................................................................................................................35计量特性....................................................................................................................45.1准确度等级..............................................................................................................45.2绝缘电阻.................................................................................................................45.3绝缘强度.................................................................................................................46校准条件....................................................................................................................46.1环境条件.................................................................................................................46.2测量标准及其他设备.............................................................................................47校准项目和校准方法................................................................................................57.1校准项目.................................................................................................................57.2校准方法.................................................................................................................58校准结果表达............................................................................................................79复校时间间隔............................................................................................................8附录A............................................................................................................................9附录B..........................................................................................................................11附录C..........................................................................................................................12附录D..........................................................................................................................15JJF(浙)1074--2012II引言JJF1071-2010《国家计量校准规范编写规则》、JJF1001-2011《通用计量术语及定义》、JJF1059《测量不确定度评定与表示》共同构成本规范修订工作的基础性系列规范。本规范是首次制定的浙江省地方计量校准规范。空气微泄漏检测仪采用高精度的差压传感器,在基准物和被测件两端同时充入相同的压力,使差压传感器处于平衡的初始状态,让仪器在依次进行充气、平衡、检测等工作过程中,如果被测件有微小泄漏,则差压传感器能探测到基准物与被测件之间的压差,最终差压信号通过一定的数学模型计算出泄漏量,并进行显示。空气微泄漏检测仪主要用于燃气具企业的燃气具密封性测试,目前,空气微泄漏检测仪在我国已有众多用户,而国内目前尚无相应的计量技术规范,该仪器的量值溯源工作处于盲区。由于没有国家校准规范,各地在校准项目、校准方法及标准器的要求等方面不能统一、不够规范和完善。因此,制订本规范。JJF(浙)1074--20123空气微泄漏检测仪校准规范1范围本规范适用于检测压力为(0~15)kPa、气体泄漏量为(0~10)ml/min的差压式空气微泄漏检测仪(以下简称检漏仪)的校准。2引用文件本规范引用了下列文件:GB/T15479-1995工业自动化仪表绝缘电阻、绝缘强度技术要求和试验方法GB/T25752-2010差压式气密检漏仪凡是注日期的引用文件,仅注日期的版本适用于本规范;凡是不注日期的引用文件,其最新版本(包括所有的修改单)适用于本规范。3术语3.1基准物master和被测物的形状、材质及容积相同且其漏率和被测物相比可以忽略的参照物体。3.2泄漏量leakagerate单位时间通过被测物内某一横截面的体积,单位为ml/min。3.3差压式检漏仪differentialpressureairleakagetester以差压传感器为感压元件检测被测物和基准物之间压力差来确定被测物是否存在泄漏和漏率大小的仪器或装置。4概述检漏仪主要适用于在测试压力小于15kPa的密闭容器里的微小气体泄漏量的检测。检漏仪采用高精度的差压传感器,在基准物和被测件两端同时充入相同的压力,使差压传感器处于平衡的初始状态,让仪器在依次进行充气、平衡、检测等工作过程中,如果被测件有微小泄漏,则差压传感器能探测到基准物与被测件之间的压差,最终差压信号通过一定的数学模型计算出泄漏量,并进行显示。检漏仪的工作原理如图1所示:JJF(浙)1074--20124压力发生装置压力发生装置减压阀测试压力传感器加压排气阀平衡阀平衡阀差压传感器基准物阀被测件阀信号处理单元显示器图1检漏仪工作原理5计量特性5.1准确度等级检漏仪的基本参数及对应的技术指标如表1所示。表1检漏仪的基本参数及对应的技术指标项目范围最大允许误差测试压力(0~15)kPa±2%泄漏量(0~10)ml/min±10%5.2绝缘电阻检漏仪绝缘电阻应符合GB/T15479-1995中4.1的规定,检漏仪电源端子与外壳间的绝缘电阻应不小于20MΩ。5.3绝缘强度检漏仪绝缘强度应符合GB/T15479-1995中4.2的规定,检漏仪电源端子与外壳间应能承受1.5kV的正弦交流电压,历时1min,无击穿、无飞弧现象。6校准条件6.1环境条件6.1.1环境温度为(20±5)℃。6.1.2环境湿度不大于80%RH。6.1.3校准设备周围应无影响读数的机械振动。6.2测量标准及其他设备校准时所需的标准仪器及设备见表2所示。JJF(浙)1074--20125表2校准用仪器设备序号仪器设备名称技术要求1数字压力计压力范围:(0~30)kPa,准确度不低于0.5级。2标准漏孔测试压力为5kPa、10kPa、15kPa的标准漏各两个,误差不大于±5%。3耐压测试仪电压范围:(0~5)kV,±5%,频率:(45~55)Hz,功率不低于0.25kW。4绝缘电阻表输出电压500V、10级。5压力发生装置可提供压力不小于0.2MPa洁净干燥空气的设备。6基准物容积为50ml、100ml和200